Automated evaluation of synovial and ascitic fluids with the Advia 2120 hematology analyzer.

用ADVIA2120血液分析仪对滑囊液和腹水进行自动评估。

Froom P,Diab A,Barak M

Abstract

The purpose of this study was to determine if the Advia 2120 hematology analyzer (Siemens Healthcare Diagnostics, Deerfield, IL) can accurately quantify both WBCs and the proportion of neutrophils from synovial and ascitic body fluids.
We analyzed 60 samples on the Advia and compared the results with manual counts and smear reviews. We also assessed the effect of adding hyaluronidase to the samples. WBC counts and the proportion of neutrophils reported by the hematology analyzer were harmonized and highly correlated with manual counts and fluid smear reviews.
The addition of hyaluronidase to the synovial fluid consistently increased the WBC counts on both manual and automated analysis (P < .001).
We conclude that the Advia hematology analyzer can be used for WBC and neutrophil counting of cells in synovial and ascitic fluids. Hyaluronidase should be added before manual or automated counting of cells in synovial fluids.

摘要

本研究的目的是确定ADVIA2120血液分析仪(伊利诺斯州迪尔菲尔德,西门子医疗诊断出品)是否可以准确地计算滑液和腹水中白细胞和中性粒细胞的比例。
我们用ADVIA分析60个样本,并把结果与人工涂片检查计数的结果进行比较。我们还评估了向样品中添加透明质酸酶的效果。血液分析仪报告的白细胞数量和中性粒细胞比例均与人工涂片检查计数的结果相差无几且高度相关。
向滑液中添加透明质酸酶均可增加人工和自动分析(P<0.001)白细胞计数结果。
我们得出结论,ADVIA血液分析仪可用于滑液和腹水的白细胞及中性粒细胞计数。透明质酸酶应在对滑液进行人工或自动计数之前加入。