EDTA Treatment of Serum Unmasks Complement-Mediated Prozone Inhibition in Human Leukocyte Antigen Antibody Testing.

人类白细胞抗原抗体检测中血清经EDTA处理后暴露补体介导前带抑制

Anani WQ,Zeevi A,Lunz JG

Abstract

Luminex-based single-antigen bead human leukocyte antigen (HLA) antibody testing is widely used to define HLA antibodies for transplant compatibility. False-negative results can occur with complement-mediated prozone inhibition. This study assessed the effect of EDTA on the assay background reactivity and fluctuations in antibody mean fluorescent intensity.
Serum specimens were retrospectively tested using Luminex-based single-antigen beads with and without EDTA. Treated and untreated serum samples were compared by two measures: changes in background reactivity and changes in HLA antibody strength after EDTA treatment.
Ten pretransplant and 48 posttransplant specimens were identified: lung (22), heart (10), kidney (21), heart/lung (two), pancreas (one), small bowel (one), and liver (one). After EDTA treatment, weak antibodies (below 2,000 mean florescent intensity) demonstrated the largest fluctuations. Newly identified HLA antibodies were seen in 16% (8/49) of class I and 26% (15/57) of class II beads. EDTA treatment did not result in false-negative reactions compared with untreated serum.
EDTA serum pretreatment mitigated complement-mediated prozone inhibition and improved accurate HLA antibody detection. The background reactivity and the false-negative rate of the assay appear unchanged.

摘要

基于液态芯片技术的单抗原磁珠人类白细胞抗原(HLA)抗体检测广泛用于定义HLA抗体移植兼容性。补体介导的前带抑制可出现假阴性。这项研究评估了EDTA在检测抗体平均荧光强度时对测定背景反应性和波动的影响。

对经EDTA处理和未经EDTA处理的基于液态芯片技术的单抗原磁珠血清标本进行回顾性检测。通过两种措施对处理和未经处理的血清样品进行比较:EDTA处理后背景反应性的变化和HLA抗体强度的变化。

对10例移植前和48例移植后的标本进行鉴定:肺(22),心(10),肾(21),心/肺(2),胰腺(1),小肠(1),肝(1)。EDTA处理后,弱抗体(平均荧光强度低于2000)表现出最大波动。新发现的HLA抗体见于16%8 / 49I类和26%15 / 57II类磁珠。与未经EDTA处理的血清比较,经EDTA处理的血清没有假阴性反应结果。

EDTA血清预处理能减轻补体介导的前带抑制,提高HLA抗体检测的准确性,而检验的背景反应性和假阴性率不变。

 