Abstract
We implemented multiple nucleic acid amplification test platforms because of the limited availability of test kits for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) during the early stages of the pandemic. Interpretation of results generated by different platforms and prioritization for testing algorithms required cross-comparison.
To compare the analytical sensitivity of 3 commercial SARS-CoV-2 molecular assays, selected samples were studied in parallel with Cobas SARS-CoV-2 test, NxTAG CoV Extended Panel, and ID NOW COVID-19 assays.
A total of 8043 SARS-CoV-2 tests performed from March 22 to April 19, 2020, were included in this study. For all 1794 positive specimens detected by the cobas SARS-CoV-2 assay, the cycle threshold (Ct) values were manually tracked and plotted to demonstrate the distribution of sample viral levels. Additionally, 50 and 63 low-positive specimens (Ct values >32) as well as 50 and 61 consecutive positive specimens by the cobas assay were tested with NxTAG and ID NOW, respectively, to estimate their relative sensitivities.
The Ct values of cobas SARS-CoV-2-positive samples were evenly distributed throughout ranges of 13.32 to 39.50 (mean, 25.06) and 13.60 to 42.49 (mean, 26.45) for ORF1 and E gene targets, respectively. NxTAG reliably detected only specimens with E gene Ct values lower than 33, and is estimated to detect 89.4% of positive specimens detected by cobas assay. ID NOW had performance variation independent of Ct value and is estimated to detect 83.5% of cobas positives.
Clinical specimens exhibit a wide range of viral burden, with a significant portion at low levels. Analytical sensitivity of testing platforms is critical for reliable detection of SARS-CoV-2 and uniform care to patients.