The Accuracy of the Sysmex UF-1000i in Urine Bacterial Detection Compared With the Standard Urine Analysis and Culture.

Erdman P,Anderson B,Zacko JC,Taylor K,Donaldson K

Abstract

- Urinary tract infections are characterized by the presence of microbial pathogens within the urinary tract. They represent one of the most common infections in hospitalized and clinic patients.
- To model the parameters of the Sysmex UF-1000i to the gold standard, urine culture, and to compare the detection of dipstick leukocyte esterase and nitrates to urine cultures and UF-1000i results.
- Data were compared from urine samples collected in sterile containers for bacterial culture and microscopic analysis. One sample was used to inoculate a 5% sheep blood agar and MacConkey agar plate using a 0.001-mL calibrated loop. The second sample was analyzed by urinalysis-associated microscopy. The media plates were investigated for growth after 18 to 24 hours of aerobic incubation at 37°C. The second sample was analyzed for bacteria and leukocytes with the Sysmex UF-1000i according to the manufacturer's guidelines. Three definitions for culture results, sensitivity, and specificity at different cutoff values were calculated for the UF-1000i.
- The negative predictive value for any positive culture in the adult population included in the study was 95.5%, and the negative predictive value for positive cultures containing growth of 100 000 or more colony-forming units was 99.3% using the Sysmex UF-1000i.
- Sysmex UF-1000i showed 98% sensitivity and 93.7% specificity with a 95.5% negative predictive value. Thus, a negative screen with the UF-1000i using defined thresholds for white blood cell counts and bacteria was likely to be a true negative, decreasing the need for presumptive antibiotics.