首页 > 期刊杂志 > 正文

Comparison of Two Commercial Matrix-Assisted Laser Desorption/Ionization-Time of Flight Mass Spectrometry (MALDI-TOF MS) Systems for Identification of Nontuberculous Mycobacteria.

Brown-Elliott BA,Fritsche TR,Olson BJ,Vasireddy S,Vasireddy R,Iakhiaeva E,Alame D,Wallace RJ,Branda JA

Abstract

This multicenter study's aim was to assess the performance of two commercially available matrix-assisted laser desorption/ionization time of flight mass spectrometry systems in identifying a challenge collection of clinically relevant nontuberculous mycobacteria (NTM).
NTM clinical isolates (n = 244) belonging to 23 species/subspecies were identified by gene sequencing and analyzed using Bruker Biotyper with Mycobacterial Library v5.0.0 and bioMérieux VITEK MS with v3.0 database.
Using the Bruker or bioMérieux systems, 92% and 95% of NTM strains, respectively, were identified at least to the complex/group level; 62% and 57%, respectively, were identified to the highest taxonomic level. Differentiation between members of Mycobacterium abscessus, M fortuitum, M mucogenicum, M avium, and M terrae complexes/groups was problematic for both systems, as was identification of M chelonae for the Bruker system.
Both systems identified most NTM isolates to the group/complex level, and many to the highest taxonomic level. Performance was comparable.