The influence of free hemoglobin and bilirubin on heparin monitoring by activated partial thromboplastin time and anti-xa assay.

Kostousov V,Nguyen K,Hundalani SG,Teruya J

Abstract

Context .- Elevated free hemoglobin (Hb) and bilirubinemia complicate extracorporeal membrane oxygenation and could affect unfractionated heparin (UH) therapy monitoring by anti-Xa assay and activated partial thromboplastin time (aPTT). Objectives .- To compare in vitro response of anti-Xa and aPTT assays to UH in samples with artificial hyperbilirubinemia and hyperhemoglobinemia and to estimate if this interference is also observed in vivo in pediatric extracorporeal membrane oxygenation. Design .- Measurement of aPTT and anti-Xa activity in plasma spiked with UH and increased concentration of free Hb and/or conjugated bilirubin. All samples with anti-Xa activity, antithrombin, free Hb, and bilirubin determination and infused dose of UH from inpatients on extracorporeal membrane oxygenation were extracted from the clinical patient database and analyzed. Results .- Each increment of free Hb by 100 mg/dL significantly shortened aPTT, whereas an increment of bilirubin by 6 mg/dL caused significant prolongation of aPTT and stepwise increase of free Hb and/or bilirubin in plasma decreased anti-Xa activity by 0.03 to 0.05 IU/mL. Extracorporeal membrane oxygenation samples with free Hb 50 mg/dL or greater had significantly lower anti-Xa activity compared with normal ones: 0.33 (0.25-0.42) versus 0.4 (0.31-0.48) IU/mL (P = .01), despite the identical UH infusion and similar antithrombin activity. Moderate increase of free Hb by 59 mg/dL was associated with absolute decrease of anti-Xa activity by 0.07 IU/mL. Conclusions .- Activated partial thromboplastin time and anti-Xa assay are affected by elevated level of free Hb and/or bilirubin in the presence of UH, and lower anti-Xa activity is noted in extracorporeal membrane oxygenation patients with elevated free Hb. Severe hemolysis and/or hyperbilirubinemia could compromise UH monitoring based on these assays.

摘要

full text

我要评论

0条评论