The usefulness of blast flags on the Sysmex XE-5000 is questionable.

Eilertsen H,Vøllestad NK,Hagve TA

Abstract

Hematology analyzers generate suspect flags that involve microscopic reviews to confirm the presence of pathologic cells. This study investigated the reliability of the blast flag in a side-by-side evaluation of 3 Sysmex XE-5000 instruments (Sysmex, Kobe, Japan). The repeatability of the Q values reported by each instrument for 10 replicates of the same blood samples was low (intraclass correlation coefficient [ICC] values, 0.62-0.74). The reproducibility of the Q values obtained by analyzing 408 samples on all 3 instruments was reasonable (ICC value, 0.85). In addition, a systematic difference was observed among the instruments in the level of reported Q values. With cutoff commonly being 100, the observed reproducibility of the blast flagging among the instruments was evaluated as poor (κ = 0.73). Based on the observed low performances, we question the usefulness of the Q value as a predictor of blasts and whether a blast flag reported by the XE-5000 is sufficient as a criterion for performing a microscopic review.