Rarok AA,Huitema MG,van der Leij MJ,van der Geld YM,Berthold H,Schmitt J,Stegeman CA,Limburg PC,Kallenberg CG
Abstract
The presence of antineutrophil cytoplasmic autoantibodies with specificity for proteinase 3 (PR3-ANCA) usually is detected by enzyme-linked immunosorbent assay (ELISA) with purified PR3 as a substrate. We studied the technical performance of direct and capture ELISA using a recombinant proteolytically inactive form of PR3 produced in the baculovirus expression system for the detection of PR3-ANCA in 114 patients with systemic vasculitis at diagnosis. We found that ELISA using recombinant PR3 produced in insect cells is a promising alternative for ELISA with native PR3. We found a correlation between tests using recombinant or native PR3, as well as correlation of the ELISA results with ANCA titers measured by the indirect immunofluorescence technique. However, the specificity for ANCA-associated vasculitis of ELISA with recombinant PR3 was lower than ELISA using native PR3. Compared with the direct assay, capture ELISA is a more sensitive method for PR3-ANCA detection, with both native and recombinant PR3, and its results depend on the monoclonal antibody used to capture the antigen.