Raboni SM,Nogueira MB,Hakim VM,Torrecilha VT,Lerner H,Tsuchiya LR

Abstract

Human rotaviruses are the most important etiologic agents of acquired diarrhea in infants and young children worldwide. Early diagnosis is essentialfor effective patient treatment. The latex agglutination (LA) assays for rotavirus diagnosis are rapid, inexpensive, and the most widely used to screen specimens. The performance of the LA Rotagen (Biokit S.A., Barcelona, Spain) was evaluated for rotavirus detection infecal samples of outpatients with acute gastroenteritis. This assay was compared with the enzyme immunoassay (EIA) EIARA (Bio-Manguinhos, Rio de Janeiro, Brazil). From January to October 2000, 285 fecal specimens were analyzed. Forty-four samples (15.4%) were reactive, 214 (75.4%) were nonreactive, and 27 (9.5%) were indeterminate by LA. All LA-positive samples were positive by EIA, and 2 LA-negative samples were positive by EIA. Of specimens indeterminate by LA, 67% were positive by EIA. The sensitivity, specificity, and accuracy of LA were 69%, 100%, and 93%, respectively. These results indicate that assay is as sensitive and specific as the EIA, and it could be applied on a large scale for screening stool specimens in suspected rotavirus diarrhea. However, the indeterminate results must be confirmed by other methods, such as EIA.